Neogen ELISA Drug Detection KitNeogen’s Triamcinolone Acetonide ELISA (Enzyme-Linked ImmunoSorbent Assay) test kit is a qualitative one-step kit designed for use as a screening device for the detection of Triamcinolone Acetonide, Budesonide, Prednisolone and other metabolites. The kit was designed for screening purposes and is intended for forensic use only.

Intended Use

For the determination of trace quantities of Triamcinolone Acetonide and/or other metabolites in human urine, blood or oral fluid. Contact a Neogen representative for information on other matrices.

Sensitivity

Compound I-50 in EIA Buffer
Triamcinolone Acetonide 0.3 ng/mL
Fluocinonide 0.3 ng/mL
Fluocinolone Acetonide 0.45 ng/mL
Amcinonide 0.7 ng/mL
Flunisolide 1.4 ng/mL
Halcinonide 1.5 ng/mL
Budesonide 5.4 ng/mL
Flurandrenolide 8 ng/mL
Diflorasone Acetate 18 ng/mL
Flumethasone 42 ng/mL
Triamcinolone 60 ng/mL
Triamcinolone Diacetate 60 ng/mL
Desoximethasone 65 ng/mL
Dexamethasone 78 ng/mL
Betamethasone 95 ng/mL
Clobestasol Propionate 100 ng/mL
Prednisolone 100 ng/mL
Beclomethasone 127 ng/mL

Specificity (Cross-Reactivity)

Compound Cross-Reactivity
Triamcinolone Acetonide 100%
Fluocinonide 100%
Fluocinolone Acetonide 67%
Amcinolone 43%
Flunisolide 26%
Halcinonide 20%
Budesonide 7%
Flurandrenolide 3.8%
Diflorasone Acetate 1.7%
Flumethasone 0.9%
Desoximethasone 0.5%
Triamcinolone 0.5%
Triamcinolone Diacetate 0.5%
Dexamethasone 0.4%
Betamethasone 0.3%
Clobetasol Propionate 0.3%
Prednisolone 0.3%
Beclomethasone 0.2%
Corticosterone 0.1%
Hydrocortisone 0.04%
Methylprednisolone 0.03%
Cortisone Acetate 0.02%
Prednisone 0.02%

Order Details

Product # Product Description
105119 Triamcinolone Acetonide, 96 well kit
105116 Triamcinolone Acetonide, 480 well kit

 

For forensic use only.

Please check with your local representative for availability of these products within your region.